Resumen: Two scintillating bolometers of LiF (33 g) and Al2O3 (50 g) at 20 mK, inside a lead shielding at the Canfranc Underground Laboratory, were irradiated with neutrons from a source of 252Cf. The analysis of nuclear recoils registered by sapphire and (n, a) captures by 6Li shows the feasibility of these cryogenic devices to measure the spectral flux of a neutron field. Data unfolding was done assuming that the spectral flux is a piecewise constant function defined on six energy groups. It can be solved by using non-negative least squares without additional assumptions on the neutron flux. The model provides consistent results with the spectra of the observed events in bolometers, giving a fast neutron flux of F(E > 0.1 MeV) = 0.20 n s-1cm-2 with a 15% uncertainty after 3 hours of live time. After our analysis, it can be concluded that nuclear recoils in sapphire are more useful than (n, a) captures in LiF for spectrometry of fast neutrons.
Idioma: Inglés
DOI: 10.1109/TNS.2016.2561280
Año: 2016
Publicado en: IEEE TRANSACTIONS ON NUCLEAR SCIENCE 63, 3 (2016), 1967-1975
ISSN: 0018-9499

Financiación: info:eu-repo/grantAgreement/ES/MICINN/FPI-CSD2009-00064
Financiación: info:eu-repo/grantAgreement/ES/MINECO-FEDER/FPA2008-03228
Financiación: info:eu-repo/grantAgreement/ES/MINECO-FEDER/FPA2011-23749
Tipo y forma: Article (PrePrint)
Área (Departamento): Física Atómica, Molecular y Nuclear (Departamento de Física Teórica)
0
Exportado de SIDERAL (2016-09-23-13:02:40)

Este artículo se encuentra en las siguientes colecciones:
articulos > articulos-por-area > fisica_atomica,_molecular_y_nuclear

Visitas


 Notice créée le 2016-09-23, modifiée le 2016-10-06


Liens externes:
Télécharger le documentPreprint
Télécharger le documentPreprint